Functions | A 3
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
51
A 3.7.3.5
Pneumatic nebulization and O
2
concentration
Remark
When the O
2
supply is not connected, the pneumatic nebulizer
cannot be started.
The nebulizer is operated with air at a set oxygen concentration (FiO
2
) ≤ 60 %
and with oxygen at a concentration > 60%.
The oxygen concentration of the inspiratory flow and the byflow are adapted
such that the set oxygen concentration (FiO
2
) at the Y-connector is guaranteed
within a tolerance range of +/- 4%.
If the pneumatic nebulizer is switched on, the Instant View window no longer
shows the measured FiO
2
value but rather the value calculated at the Y-connector.
In this case the colour coding of the display field is hidden.
A 3.8
Hygiene function
After selecting the hygiene function via the “Functions” button, you have the fol-
lowing options:
• you can activate and deactivate the currently used accessories with “On” and
“Off”, respectively;
• you can view the remaining times of use of the accessories;
•
you can confirm the replacement of an accessory.
The following components are monitored:
• Breathing circuit
• Heated breathing circuit
•
turbine filter
• valve bar
• nebulizer
• suction system
Remark
Components which were set to “Off” at the configuration level
are not visible in this screen.
When the period of service of an accessory has expired, the alarm window will
display a message.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...