Functions | A 3
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
67
Remark
By touching the “On” button, you activate the device. After activa-
tion, please bear the following information in mind:
•
The first measured value and the corresponding waveform will
not be displayed before the end of the warm-up phase, which
may take up to 10 seconds.
•
The accuracy of the measurement is reduced in the first 10
minutes because the sensor is still warming up. During this
warm-up phase, you will see the alarm message “LeoLyzer sen-
sor temperature”.
•
In the first 10 minutes, the accuracy of the measurement is very
much dependent on the device temperature. Low temperatures
of the device lead to strong fluctuations of the measured values.
Stop the multi-gas measurement by activating the “Off” button in the LeoLyzer
Workbench.
The concentration of the selected gas / gases is measured continuously. The meas-
ured values will be displayed only if previously configured in one of the views (for
details, please refer to chapter 5 General information in the Basic Manual of the
instructions for use).
The measured values can be displayed in the following formats:
• Data Table
• Short Trend Graph
• Trend Table
• Trend Graph
Interrupting the ventilation (“Standby”) also stops the multi-gas measurement.
If ventilation continues with the “Same patient”, the set alarm limits for the mul-
ti-gas concentration are retained.
When the “New patient” function is selected, the alarm limits for the multi-gas
concentration are set to the default values.
Remark
LeoLyzer auto-calibrates continuously. The user is not required to
perform calibrations.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...