A 3 | Functions
68
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
A 3.12
SpO
2
sensor
A 3.12.1
Connecting the SpO
2
sensor (with adapter)
Warning
First connect the SpO
2
finger sensor adapter to the device (power
supply) before you allow it to come into contact with the patient.
Warning
Be sure to connect the SpO
2
finger sensor adapter to one of the
BF interfaces on the back of the device.
Remark
The SpO
2
finger sensor can be used within the patient environ
-
ment.
SpO
2
sensor adapter
Back
SpO
2
finger sensor
Select the finger sensor size suitable for the patient. Connect the SpO
2
finger
sensor to the configured BF interface on the back of the ventilator using the SpO
2
adapter.
Attach the SpO
2
finger sensor to the patient’s finger.
A 3.12.1.1
SpO
2
Measurement
For details of the SpO
2
measurement, refer to the Instructions for Use, Supple-
ment D – Pulse oximetry and LeoClac.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...