Functions | A 3
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
71
Remark
Using the AnaConDa function requires permanent monitoring of
the anaesthetic gas concentration by means of an appropriate gas
monitor.
Warning
Furthermore, observe and follow the instructions given in the
documents supplied with each system component.
A 3.13.2.2
Patient safety
elisa 300 / elisa 500 with AnaConDa may only be operated by qualified medical
staff in order to guarantee an adequate intervention in the event of a device
malfunction.
Warning
You must undergo training for each system component before you
are allowed to use the system for patient treatment. Please con
-
tact your sales partner for information about training programs
for AnaConDa or AnaConDa-S in combination with elisa 300 /
elisa 500.
Warning
If the gas or power supply fails, the ventilator may not be able to
maintain the ventilation function.
In this situation, perform emergency ventilation immediately by
following these steps:
• Disconnect AnaConDa or AnaConDa-S from the patient tube
or tracheostomy tube
• connect a resuscitation bag to the patient tube or tracheosto-
my tube.
•
Ventilate the patient manually, if necessary with oxygen supply
and PEEP valve.
• Stop administering the anaesthetic agent with the syringe pump.
• Choose an alternative method if a sedative needs to be admin-
istered.
A 3.13.3
Technical requirements
Remark
Before you can use the AnaConDa function, it must be enabled at
the configuration level of elisa 300 / elisa 500.
The following additional accessories are required for use of the AnaConDa func-
tion of elisa 300 / elisa 500:
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...