Ventilation modes and parameters | A 5
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
105
A 5.5.5
Mandatory BiLevel
Synchronised, pressure-controlled ventilation mode, allowing the patient to
breathe spontaneously at the upper level and initiating a mandatory breath for
each trigger variable.
In contrast to the conventional BiLevel mode, the ventilator performs the entire
work of breathing. Consequently, BiLevel breaths are time-cycled via the set res-
piratory rate and also initiated via the trigger function.
This mode does not allow additional pressure support, nor does it allow man-
datory inspiration time to be shortened in response to an expiratory trigger
variable, which would relieve the patient significantly. The tidal volume delivered
is determined by the difference between the lower pressure level (PEEP) and the
upper pressure level (Pinsp), and is dependent upon lung elasticity (compliance).