Ventilation modes and parameters | A 5
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
115
Warning
Adapt the backup ventilation to the patient’s individual ventilation
situation.
Warning
Proportional PSV is only available for invasive ventilation. Leakage
may lead to over-compensation of the flow resistances.
Warning
Check that the conditions for using the proportional PSV venti
-
lation mode, such as an adequate respiratory drive, no hiccup, or
absence of an inhomogeneous breathing pattern are met.
Warning
Consider the contraindications for the proportional PSV ventila
-
tion mode, such as central respiratory depression, quick change
of respiratory mechanics, Cheyne-Stokes respiration, manifest or
suspected elevated intracranial pressure.
Warning
As a measure to prevent pressure peaks resulting from potential
overcompensation, both the upper pressure limit (Pmax) and the
alarm limits need to be adapted to the patient condition.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...