A 5 | Ventilation modes and parameters
122
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
A 5.5.19
Optional VCV
Synchronised, volume-controlled ventilation, allowing the patient to breathe
spontaneously at PEEP level or with the set pressure support. If the measured
minute volume is above the setting (Rate x Vt), the patient is able to breathe
spontaneously. If the minute volume drops below the setting, volume-controlled
breaths are triggered.
Optional VCV combines the advantages of volume-controlled ventilation with the
possibility of spontaneous breathing at PEEP level with a set pressure support. In
contrast to conventional SIMV ventilation, the user defines a minimum minute
volume by setting the respiratory rate and tidal volume. No mandatory breaths
are initiated if the minute volume can be achieved by the patient's spontaneous
breathing.