Instructions for Use Supplement B – CO
2
Sensor LeoCap
|
Rev. 04
15
Remark
Calibration of the LeoCap sensor is an automatic routine function.
4.5
Removing the sensor from the patient
Caution
When removing the sensor, do not hold only the cable or only
the sensor. The cable may break or the sensor may be damaged by
excessive force.
4.5.1
Separating the sensor from the airway adapter
Hold the adapter in the breathing circuit with one hand and the sensor with the
other hand. Lift off the sensor.
4.5.2
Removing the airway adapter from the breathing circuit
Disconnect the airway adapter from the breathing circuit without exerting ex-
cessive force.
4.6
Checks before and after use
After unpacking a new sensor, handle it with care because static electricity, water
or chemical solutions may cause damage.
Before each use, check the sensor and the external surfaces of the reusable air-
way adapter (especially the optical elements) for contamination. Contamination
and moisture can reduce the measuring accuracy of the CO
2
sensor. If necessary
wipe the windows clean with a moist cloth.
Also make sure that the connectors and cables are intact.
A damaged airway adapter must be replaced with a new one.
If necessary, clean and disinfect the CO
2
sensor and the airway adapter according
to chapter 5; all surfaces must be dry and clean.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...