Sensor LeoCap
|
Rev. 04
9
Technical data
Technical data
Sensor type
CO
2
mainstream sensor
Functional principle
Non-dispersive infrared sensor (NDIR)
Initialization time
20 s (at an ambient temperature of 25 °C)
CO
2
measuring range 0 – 20 vol% (0 – 150 mmHg)
CO
2
measuring
accuracy
±(0.2 vol% + 2 % of reading)* or
±(1.5 mmHg + 2 % of reading)*
CO
2
measurement
resolution
0.1 vol% or
0.1 mmHg
Response time
approx. 10 ms
Respiratory rate
0 – 200 breaths per minute
Accuracy of the res-
piratory rate meas-
urement
±1 breath
Calibration
No user calibration required
Airway adapter
for adults / children (reusable),
for children / neonates (reusable)
Material: polysulphone
Dead space volume:
<5 ml (adult/pediatric),
<1 ml (pediatric/neonate)
Sterilization:
ethylene oxide/autoclaving up to 100 times
Compliance with ISO 5356-1:2015
Compensation
Atmospheric pressure: 390 – 900 mmHg
Adjustable compensation for O2, N2O and anaesthetic
gas
Electrotechnical data
Connector
RS232
4-pin Lemo Redel
Power
Voltage: 5.0 V ± 5%
Power consumption: 1.3 W;
less than 2 W during warm-up
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...