26
Instructions for Use Supplement B – CO
2
Sensor LeoCap
|
Rev. 04
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic
environment –
guidance
Conducted RF
IEC 61000-4-
6:2013
3 Vrms
150 kHz to 80
MHz
3 Vrms
Portable and mobile RF com-
munications equipment should
be used no closer to any part
of the device, including cables,
than the recommended sepa-
ration distance calculated from
the equation applicable to the
frequency of the transmitter.
Recommended separation
distance
d = 1.2√P
Radiated RF
IEC 61000-4-
3:2008
3 V/m
80 MHz to 2.7
GHz
3 V/m
d = 1.2√P
(80 MHz to 800 MHz)
d = 2.3√P
(800 MHz to 2.7 GHz),
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and
dis the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey (a)
should be less than the com-
pliance level in each frequency
range. (b)
Interference may occur in the
vicinity of equipment marked
with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...