Instructions for Use Supplement C – Multi-Gas Sensor LeoLyzer
|
Rev. 02
7
Front view
F
G
O-rings
LED display
G
F
Rear view
H
J
Interface cable connector
Sample gas outlet
J
H
2.4
Theory of operation
The device allows the continuous measurement of CO
2
, isoflurane and sevoflu
-
rane gas concentrations in the patient's respiratory air flow by infrared spectros
-
copy.
The method is based on the measurement of infrared light absorption at wave-
lengths of 4.2 μm, 7.85 μm, and 8.3 μm. The measured amount of light that passes
through the gas to the photo sensor serves as the calculation basis. The con-
centration of CO
2
and anaesthetic gas is calculated from the partial pressure at
atmospheric pressure conditions.