8
Instructions for Use Supplement C – Multi-Gas Sensor LeoLyzer
|
Rev. 02
The device is connected to the breathing circuit using the airway adapter and the
sampling line.
An infrared light source and a photo sensor are integrated in the device. The gas
mixture to be analysed is conducted through a measuring cell. The light emitted
by the infrared light source reaches the photo sensor after passing through the
measuring cell. Thus, the device measures the degree of absorption of infrared
light during its passage through the gas stream.
3
General safety instructions
These instructions for use are an integral part of the delivery. When the device
changes ownership, the manual must also be handed over. Information contained
in this document can be changed without notice.
The device may only be operated by qualified medical staff (physicians, certified
nursing staff, nursing staff specializing in intensive care medicine, respiratory ther-
apists) in order to guarantee an adequate intervention in the event of a device
malfunction.
Before working with this device, carefully read these instructions for use as well
as the Basic Instructions for Use and Supplement A of the intensive care ventila-
tor used. Improper use of the device can cause malfunction, loss of performance
and it can create a hazard for the patient.
If malfunctions or unstable operation of the device occur, refer to the trouble-
shooting tips in section 4.11.
The following signs and symbols are used to identify important information in
these instructions for use:
Refers to dangerous situations which may occur when failing to
observe the present instructions for use.
Warning
“Warning” refers to a situation which may cause injuries to the
user or the patient.
Caution
“Caution” refers to a situation which may cause damage to the
equipment.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...