Instructions for Use Supplement C – Multi-Gas Sensor LeoLyzer
|
Rev. 02
9
Reusable consumables have a limited service life. Various operating conditions as
well as cleaning, disinfection and sterilization procedures can adversely affect the
material and reduce the service life. If signs of deterioration (cracks, deformation,
discolouration, paint peeling off, etc.) are detected, you need to replace the con-
sumables.
3.1
Safety information
Warning
If an emergency occurs during operation (fire, short circuit, etc.),
immediately disconnect the device from the patient and continue
monitoring with another device.
Warning
Use the device only with the specified intensive care ventilators.
Using devices not explicitly mentioned may cause skin burns.
Warning
Do not pull or bend the interface cable of the device. Failure to
follow these instructions may cause cable discontinuity, short cir-
cuits, failed measurements and skin burns on the patient due to
elevated device temperatures resulting from the short circuit in
the cable. Replace defective devices with new ones.
Warning
Do not let the patient bite on the interface cable or sampling line.
This may damage to the device and harm the patient.
Warning
Keep magnetic objects away from the device. Magnetic objects
may distort the waveform and cause incorrect readings.
Warning
The device is not defibrillation-proof. It is mandatory to remove
the device from the patient during defibrillation.
Warning
Empty the water trap reservoir when it is half full. It should never
be filled to the marked maximum fill level. Remember that liquid
getting into the measuring path of the device will lead to malfunc-
tion.
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...