Instructions for Use Supplement C – Multi-Gas Sensor LeoLyzer
|
Rev. 02
Remark
The following factors may impair the CO
2
, ISO and SEV measuring
accuracy:
• mechanical damage of the device
• cyclical pressure peaks of up to 10 kPa (100 cmH
2
O)
• leakage or internal venting of the sampled gas
Remark
The respiratory rate (RR) has no influence on the measuring accu
-
racy of the CO
2
, ISO and SEV concentration within the specified
RR measuring range and at an I:E ratio of 1:1.
Remark
There is no drift in the CO
2
, ISO and SEV measuring accuracy for
at least 6 hours.
10
Guidance and manufacturer's declaration
Electromagnetic environment:
Refer to the appropriate chapter in the instructions for use of the connected
intensive care ventilator.
Electromagnetic emissions
Emission
tests
Compli-
ance
Electromagnetic
environment – guidance
RF emissions
CISPR 11:2009
Group 1
The device uses RF energy only for its
internal function. Therefore, its RF emis-
sions are very low and are not likely to
cause any interference in nearby electronic
device.
RF emissions
CISPR 11:2009
Class A
The device can be used in industrial areas
and hospitals (CISPR 11 class A). If used in
a residential environment (for which CISPR
11 class B is normally required), this device
might not offer adequate protection to
radio-frequency communication services.
The user might need to take mitigation
measures, such as relocating or re-orient-
ing the device.
Harmonic emis-
sions
IEC 61000-3-
2:2005
Class A
Voltage fluctu
-
ations/ flicker
emissions
IEC 61000-3-3:95
+A1:2001
Complies
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...