Instructions for Use Supplement D – Pulse Oximetry and LeoClac
|
Rev. 01
13
5.2
Safety instructions
Warning
Observe the instructions and warning of the pulse oximetry chap-
ter. Do not employ the automatic closed-loop control function if
the clinical situation of the ventilated patient suggests a reduced
significance or accuracy of the SpO
2
measurement.
Warning
Activate the LeoClac function only when the patient condition
warrants the continued acquisition of the SpO
2
signal. Assess po-
tential contraindications such as restlessness or reactive psychosis
of the patient.
Warning
When using the LeoClac function, the user is required to ensure
the plausibility of the SpO
2
measurement by means of additional,
independent means (e.g. additional SpO
2
monitoring with a vital
signs monitor, regular blood gas analysis, ...).
Warning
Make sure that the SpO
2
measurement is unimpeded. Do not apply
the sensor on the same arm used for non-invasive blood pressure
measurement or IABP therapy.
Warning
LeoClac may only be used in patients with sufficient minimum ven
-
tilation.
5.3
Function
• Select and activate the SpO
2
function (see also chapter 5)
• Select the LeoClac function
•
Enter the SpO
2
target range
• Adapt the alarm limits for “O
2
ctrl. min/max” to the clinical situation
• Adapt the alarm limits for “+SpO
2
ctrl. min/max” to the clinical situation
• Start the LeoClac function (“On”)
Summary of Contents for elisa 300
Page 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Page 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Page 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Page 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Page 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Page 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Page 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Page 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...