A 3 | Functions
62
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
Remark
For display of the volumetric capnography information, a special
template must be selected on the configuration level. Please con
-
tact the equipment officer of your institution or an authorized
service agent.
Remark
VCO
2
reflects the measured CO
2
elimination which may be dif-
ferent from the CO
2
production because it is influenced by the
clinical situation.
A 3.10.10
Alarm limits
The following alarm limits are available for an adequate monitoring of the CO
2
measurement:
etCO
2
max
Upper alarm limit for the expiratory CO
2
concentration
etCO
2
min
Lower alarm limit for the expiratory CO
2
concentration
inCO
2
max
Upper alarm limit for the inspiratory CO
2
concentration