DESINFECTING THE MICRO-HEADS AND MICRO-NOZZLES
The motorised treatment heads (Alliance 80, Alliance 50 and TR30) need to be used with
an Endermowear suit. Non-motorised treatment heads (micro-nozzles and micro-heads)
can be used directly on the skin in specific cases.
In these cases, the heads need to be disinfected after each use.
1. Use the maintenance procedure described above.
2. Soak the rollers, micro-heads, disassembly tool and micro-nozzles in a disinfectant for
12 minutes at 20 °C, as recommend on the disinfectant packaging.
3. Carefully rinse the flap and the treatment chamber with sterile or drinking water for
at least one minute using a large volume of water (approximately eight litres). Repeat
twice for a total of three rinses.
4. Dry the parts.
5. Pre-clean the storage drawer using LPG
®
wipes then place the head in it.
Use a disinfectant whose active ingredient is ortho-phthalaldehyde (OPA). Before using the disinfectant, read and follow
the recommendations, contraindications and warnings associated with this product. Refer to the instructions for using
this solution. All the procedures described in this section must be carried out with the machine turned off and the power
cord unplugged. - Do not use corrosive products such as acetone, trichloroethylene, or rubbing alcohol, nor abrasive
sponges.
≥