2 3
TECHNIC
AL
SPECIFIC
A
TIONS
6
→
TECHNICAL SPECIFICATIONS
Dimensions; (LxWxH): . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 780 x 680 x1660 mm
Net weight: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76.1 kg
Maximum set depression: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69 kPa (690 mbar)
Cooling: . . . . . . . . . . . . . . . . . . . . . . . . . . . By mechanical ventilation incorporated in the pump
Protection index: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IP20
Electrical protection class: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Operating temperature: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . +10 °C to +30 °C
Storage temperature: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -20 °C to +70 °C
Electrical features: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100-240 V / 50-60 Hz / 710-670 W
Operating environment:
Ambient temperature: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 °C to 30 °C for normal operation.
Ambient relative humidity: . . . . . . . . . . . . . . . . . . . . . . . . . . .30% to 75% without condensation.
Atmospheric pressure: . . . . . . . . . . . . . . . . . . . . . . . . . . no significant influence for operation.
Max altitude: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2,500 m
Unit fitted with patented treatment heads.
1912
The Cellu M6
®
Integral
i
is marked as a medical device by virtue of Annex II of
regulation 93/42/EEC (applicable standards IEC 60601-1 Ed3 and related standards).
ELECTROMAGNETIC COMPATIBILITY
For more information about electromagnetic compatibility, refer to the “Electromagnetic
Compatibility” appendix.