APPENDIX
: ELEC
TR
OMA
GNETIC
C
OMP
A
TIBILIT
Y
10
Conducted RF disturbances
IEC 61000-4-6
3 V 0.15 MHz – 80 MHz
6 V in ISM bands and
amateur radio bands
between 0.15 MHz and
80 MHz
80% AM at 1 kHz
3 V 0.15 MHz – 80 MHz
6 V in ISM bands and
amateur radio bands
between 0.15 MHz and
80 MHz
80% AM at 1 kHz
3 V 0.15 MHz – 80 MHz
6 V in ISM bands and amateur radio bands
between 0.15 MHz and 80 MHz
80% AM at 1 kHz
Power frequency magnetic
field
IEC 61000-4-8
30 A/m
30 A/m
30 A/m
Voltage Dips and
Interruptions:
IEC 61000-4-11
0% UT; 0.5 cycle
A 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0% UT; 1 cycle at 0°
70% UT; 25/30 cycles at 0°
0% UT; 250/300 cycles
0% UT; 0.5 cycle
A 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0% UT; 1 cycle at 0°
70% UT; 25/30 cycles at 0°
0% UT; 250/300 cycles
0% UT; 0.5 cycle
A 0°, 45°, 90°, 135°, 180°, 225°, 270° end 315°
0% UT; 1 cycle at 0°
70% UT; 25/30 cycles at 0°
0% UT; 250/300 cycles
Your CELLU M6
®
Integral i device requires special care concerning the EMC; it must be
installed and serviced according to the information provided in this user guide.
Portable and mobile RF communication devices must not be used within 30 cm of your device;
they can cause undesired operation.
The use of treatment heads other than those provided by LPG may result in increased
emissions or decreased immunity of the device.
Your CELLU M6
®
integral i device should not be used adjacent to or stacked with other medical
devices.
The Cellu M6
®
Integral i device does not manage essential performances.
Interference may occur near equipment marked with the following symbol:
→
APPENDIX: ELECTROMAGNETIC COMPATIBILITY