30
31
4 service sheets: structural components
service sheets: structural components
4.6 rePlacing tHe toP Structure
Safety:
For safety reasons, the CELLU M6 safety operator or technician should not continue
when the machine is powered on or in 230V areas.
Unplug the power cord from the outlet.
required materials:
1 x T20 torx key, 1 x T30 torx key
operating time:
45 min.
1)
Follow the instructions from step 1 to 2:
4.3 rePlacing a Head HoSe
3)
Unlock the ring by lining up the
two ¦ symbols, and then pull the
ring to disconnect the hose.
(Disconnecting: See note at the start
of the chapter.)
6)
Disconnect the cable
from the panel.
4)
Remove the hose from the
movable arm, with its guide.
7)
Unscrew the four screws (T20
Torx) from the panel and remove
the entire panel.
2)
Open the filter drawer and
pull the lid from the frame arm
to remove it. Repeat the same
operation on the second arm.
8)
Unscrew the large
screws (T20 Torx) from the
panel and set them aside.
5)
Unscrew the two screws
from the rear cover of the
panel.
tilt the panel
down
to remove the cover.
cont .
>
9)
Unscrew (T30 Torx)
and remove the both
movable arms.
10)
Unscrew the
two
bases
(right and left)
from the frame.
4 service sheets: structural components
service sheets: structural components
4.5 rePlacing tHe Panel adjuStMent Brake
Safety:
For safety reasons, the CELLU M6 safety operator or technician should not continue
when the machine is powered on or in 230V areas.
Unplug the power cord from the outlet.
required materials:
1 x T20 torx key.
operating time:
20 min.
note:
adjust the hose’s position
so that it runs along the joint of
the moving arm when in its
lower position.
1)
Open the filter drawer and pull
the lid for the frame arm to remove
it. Remove the top two prongs.
Repeat the same operation on the
second arm.
4)
Unscrew the four screws (T20
Torx) from the panel and remove
the entire panel.
5)
Unscrew the large screws
(T20 Torx) from the panel and
set them aside.
3)
Disconnect the cable
from the panel.
2)
Unscrew the two screws
from the rear cover of the
panel. Tilt the panel down to
remove the cover.
reassembly:
Perform the
operations in reverse to reassemble
the new assembly.