36
37
4 service sheets: electronic and electrical components
service sheets: electronic and electrical components
1)
Follow the instructions from step 1 to 3:
4.13 rePlacing tHe PoWer SuPPly
2)
Disconnect all of the
cables, marking their
respective positions.
connector Positions
reassembly:
Perform the operations in reverse to reassemble the new assembly.
3)
Disconnect the host from the
suction sensor exhaust.
Unscrew the four nuts (3mm
key) and remove the board.
4.14 rePlacing tHe regulation Board
Safety:
For safety reasons, the CELLU M6 safety operator or technician should not continue
when the machine is powered on or in 230V areas.
Unplug the power cord from the outlet.
required materials:
1 x T20 torx key, 1 x key with a 3mm fork or tube, 1 x 3mm hex key.
operating time:
15 min.
head 2
light
strips
pc panel
multiple
voltage power
sensor 1
sensor 2
evp
evs
fan
nc
4 service sheets: electronic and electrical components
service sheets: electronic and electrical components
1)
Unscrew the screws
attached to the cover under
the wheel legs (3mm hex key,
depending on the version).
Remove the wheel covers.
4)
Unscrew the four retaining
screws (T20 Torx), and then
disconnect:
The power cable from the
•
control card (1)
Both pump power cables
•
(2), marking their respective
positions, and unscrewing the
locks (3mm flat screwdriver)
Disconnect the pumps from
•
the grounding (3) (If needed,
depending on the version).
2)
Unscrew the locking
screw on the top of the
power socket.
5)
Rotate the plate to the
left, being careful not to
pull on the strands of cable,
and then unscrew the four
retaining screws (T20 Torx)
from the power unit at the
back of the plate.
3)
Unclick the top/bottom
covers by firmly pulling down.
4.13 rePlacing tHe uniVerSal PoWer SuPPly
Safety:
For safety reasons, the CELLU M6 safety operator or technician should not continue
when the machine is powered on or in 230V areas.
Unplug the power cord from the outlet.
required materials:
1 x T20 torx key, 1 x 3mm flat screwdriver, 1 x 3mm hex key.
operating time:
15 min.
reassembly:
Perform the operations in reverse to reassemble the new assembly.
1
2
3
head 1