configuration and maintenance menus
6 configuration and maintenance menus
>
6.9c Functional teStS taB (cont.)
Button box screen (cont.)
regulation
This table displays the pressure in mBar (Vacuum column) calculated for each sensor.
The ‘Differential Vacuum’ column displays
the differential pressure each second and
corresponds to residual leaks.
temperatures
This table displays the temperatures recorded by different sensors.
The temperature displayed beneath
‘Controller t°’ corresponds to the regulation
board temperature.
The temperatures displayed in the table
correspond to those measured on various
units of power supply hardware.
Button
action
TIMER:
A timer can be launched as required. It displays time in hours / minutes / seconds
Start Count
Start timer
Stop Count
Stop timer
PUMPS MANAGEMENT
Pump 1 On
Start Pump 1
Pump 2 On
Start Pump 2
Pump 1 & 2 On
Start Pump 1 and 2
OFF
Stop Pump 1 and 2
ADJUSTING THE LIGHT BAND
ON
Switch on the light band in red
OFF
Switch off the light band
ADJUSTING THE FAN
ON (100%)
Fan on at full speed (3600 rpm ±10%)
OFF (0%)
Adjust the fan speed to 0% (900 rpm ± 10%)
cont .
>
configuration and maintenance menus
6 configuration and maintenance menus
Button
action
SELECTION SOLENOID VALVE ADJUSTMENT:
These solenoid valves allow selection of Arm 1, Arm 2 or both
Open all
Open all Selection Solenoid valves
EVS1 Open
Open Selection Solenoid valve 1
EVS2 Open
Open Selection Solenoid valve 2
Close All
Close all Selection Solenoid valves
EVS1 Close
Close Selection Solenoid valve 1
EVS2 Close
Close Selection Solenoid valve 2
PRESSURE SOLENOID VALVE ADJUSTMENT:
Once a regulation is launched, all solenoid valves are checked by the regulation board
Open all
Open all The pump and atmospheric pressure
solenoid valves
EVP1 Open
Open the Pump Solenoid valve 1
EVP2 Open
Open the Pump Solenoid valve 2
EVPA Open
Open the Atmospheric Pressure solenoid valve
Close All
Close all solenoid valves
EVP1 Close
Close Pump solenoid valve 1
EVP2 Close
Close Pump Solenoid valve 2
EVPA Close
Open the Atmospheric Pressure solenoid valve
CyCLE RATE ADJUSTMENT
-
Decrease the cycle rate by 5%
+
Increase the cycle rate by 5%
VACUUM ADJUSTMENT
-
Decrease the pressure by 50 mBar
+
Increase the pressure by 50 mBar
FREQUENCy ADJUSTMENT
-
Decrease the frequency by 0.1 Hz
+
Increase the frequency by 0.1 Hz
>
6.9c Functional teStS taB (cont.)
Button box screen (cont.)
cont .
>