DRUCK
TEMPERATUR
&
Leitenberger GmbH •
Bahnhofstr. 33 • D-72138 Kirchentellinsfurt • GERMANY
Tel. +49 (0) 7121-90920-0 • Fax +49 (0) 7121-90920-99 •
-Info@Leitenberger.de •
• www.druck-temperatur.de
DT
www.
-Cal.net
LR
MANUAL
LHM - Handmessgerät für Druck, Kra/Gewicht und Drehmoment
LHM - handheld for pressure, force/weight and torque
54
Δ1
Δ2
P1
P2
P3
First
PEAK
Max.
PEAK
measur
ed v
alue
me
P3 is the PEAK
value if the „First
PEAK“ detecon
is disabled.
If „First PEAK“ detecon is enabled:
P1 is not validated as First Peak because the
decreasing measured value 1 aer the peak
Δ
is lower than the threshold (%FS) set.
P2 is validated as First Peak because the
decreasing measured value 2 aer the peak
Δ
is greater than the threshold (%FS) set.
•
Enable - for
with built-in absolute pressure sensor
Zero Key
LHM
LR-
Cal
Normally the measure for these kind of sensors is absolute pressure and it is not necessary
to
the value, the
funcon is disabled.
ZERO
ZERO
Any way, in some applicaons it is usefull to have the possibility to ZERO the current value.
The parameter
when set to
allows this funcon.
Zero Enable
Yes
•
TIME AVERAGE
This funcon allows you to calculate for long periods of me (up to 60 seconds) the average
value of the measurement. The average value is updated every second. When enabled, you
can acvate the feature in the main page using the F2 key. When this mode is acvated, the
PEAK mode is disabled and vice versa.
On this page you can enable
/
) the funcon and set the me with respect to which
(YES NO
the average is calculated.
7.8 Lang.
(Language)
Do not change the operang language (English).
7.9 RF Communicat.
(WIRELESS communicaon parameters)
On this page you can change some parameters that affect to the oponal RF communicaon
(WIRELESS). See next page.