In the main menu, select item „Service“ and enter via arrow-keys the password
for
1234
posive values, or password
for negave values and confirm with ENTER.
1235
If necessary press ZERO bu on to acvate the ZERO funcon.
Choose the calibraon points Px (x from 1 to 5) and confirm with ENTER.
Load the
exactly with referring pressure/force/torque and compare the true value
LHM
LR-
Cal
with the indicated value. You may correct the indicated value with the arrow-keys and confirm
with ENTER.
Make the up to 5 comparisons/calibraons. If you have finished the linearisaon, select „Quit“.
13. Dismounng, return and disposal
WARNING!
Residual media in the dismounted
can result in a risk to
LHM
(sensor)
LR-
Cal
persons, the environment and the equipment. Take sufficient precauonary
measures.
13.1 Dismounng
Only disconnect the test and calibraon installaons once the system has been unloaded!
13.2 Return
WARNING!
Strictly observe the following when shipping the
LHM equipment
LR-
Cal
All instruments, delivered to DRUCK & TEMPERATUR Leitenberger GmbH, must
be free from any kind of hazardous substances (acids, bases, soluons, etc.).
When returning the
use the original packaging or a suitable transport package.
LHM
,
LR-
Cal
To avoid damage:
1. Wrap the
in an anstac plasc-film.
LHM
LR-
Cal
2. Place the
along with the shock-absorbent materila, in the packaging.
LHM
LR-
Cal
Place shock-absorbent material evenly on all sides of the transport packaging.
3. If possible, place a bag, containing a desiccant, inside the packaging.
4. Label the shipment as transport of a highly sensive measuring instrument.