LSI LASTEM M-Log – User’s manual
30
During auto-recognition function could be pointed out some errors; here below the codes and their
relevant meanings:
1
GR_CISS_Error
In the instrument the library of sensors includes the
description of sensor or non-valid measurement
2
GR_EmptyLibrary
The instrument doesn’t include a valid library of
sensors
3
GR_ParamNotFound
The measurement doesn’t include the placeholder
for demanded parameter
4
GR_UndefinedProbe
The connected sensor hasn’t been defined in library
of instrument
5
GR_UndefinedMeasure
The connected sensor generates a measurement that
hasn’t been defined in library of instrument
6
GR_DepMeasureNotFound
One calculated measurement requires one
dependent measurement (acquired or calculated by
same or other sensors) that hasn’t been generated
7
GR_DepMeasureMarkerNotFound
The tag for dependent measurements, included into
measurement definition, hasn’t been found
8
GR_MeasureUpdateRateNotFound
The tag for the update rate, included into
measurement definition, hasn’t been found
9
GR_SensorCalibrMarkerNotFound
The tag for calibration value, included into
measurement definition, hasn’t been found
10
GR_UndefinedProbeCalibrParam
One connected sensor requires the definition of
calibration value that hasn’t been defined in
standard quantities (*)
11
GR_UpdateActiveConfigError
Memory writing error during update of instrument
current configuration
(*)
Algorithm for search of parameters for sensor calibration: if there’s only one defined parameter
it’s always used for every input (also more than one); if there’re several defined parameters, they
must correspond with specified input; in other cases, error. The calibration parameters can be
defined at the end of list of standard quantities.
The function of probes auto-recognition is a local setup, in other words the configuration is set-up
automatically and just in the instrument where this function is required. In case there’s a network of
M-Log dataloggers, it’s necessary the use of 3DOM software in order to update the configuration of
master instrument and acquire properly all quantities coming from other acquirers by radio.