32
Maximum Voltage between
any Terminal and Earth Ground
1000 V DC, 750 V AC
Fuse Protection
mA or
μ
A
2 A, 600 V high energy fast fuse
A
15 A, 600 V high energy fast fuse
Display (LCD)
digital
Counts: 2000, updates 3/sec
Operating Temperature
0 °C to 40 °C (32 °F to 104 °F)
storage Temperature
- 20 °C to 60 °C (-4 °F to 140 °F)
Temperature Coefficient
0.05 x (specified accuracy) °C
(<18 °C or >28 °C; <64 °F or > 82 °F)
Relative Humidity
0 % to 80 % (0 °C to 35 °C; 32 °F to 95 °F)
0 % to 70 % ( 35 °C to 55 °C; 95 °F to 131 °F)
Battery Type
9 V, NEDA 1604 or 6F22 or 006P
Battery Life
200 hrs typical with alkaline
size (H x W x L)
Meter only
3.2 x 8.6 x 18.7 cm (1.25 x 3.39 x 7.35 in)
With Holster & Stand
4.7 x 10.0 x 20.3 cm (1.85 x 3.94 x 7.99 in)
Weight
Meter only
332 g (11.7oz)
With Holster & Stand
570 g (20.1 oz)
safety
Designed to Protection Class II per IEC 348,
CSA C22.2 NO. 231, ISA-DS82, and UL 244
– Fuse Replacement
Referring to
figure 4, page 46,
use the following procedure
to examine or replace the meter‘s fuses:
1. Perform steps 1 through 3 of the battery replacement
procedure.
2. Remove the defective fuse by gently prying one end
of the fuse loose and sliding the fuse out of the fuse
bracket.
3. Install a new fuse of the same size and rating.
Make sure the new fuse is centered in the fuse holder.
4. Replace the case top, ensuring that the battery leads
are properly dressed and the two snaps on the case top
(at the end near the LCD) are engaged. Reinstall the
three screws.
– service
If the meter fails, check the battery and fuse(s) and replace
as needed. lf the meter still does not work properly, review
this manual to make sure you are operating it correctly. If
the meter still malfunctions, pack it securely in its original
shipping container and forward it, postage paid, to the
nearest distributor. lnclude a description of the maIfunction.
Finest assumes No responsibility for damage in transit. A
meter under warranty will be promptly repaired or replaced
and returned at no charge. If the warranty has laped, the
meter will be repaired and returned for a fixed free. Contact
the nearest distributor for information and prices.
7. REPLACEMENT PARTs
NOTE
Only genuine spare parts can be used
when performing service work on
the device!
Replacemant parts are shown in
figure 5, page 47
and lis-
ted in
table 3, page 32
. To order replacement parts contact
the nearest distributor.
Table 3.
Replacement parts
Item
Description
Part number
ET 200* Fuse 2 A, 600 V RMS
120421
ET 201
Fuse 15 A, 600 V RMS
120422
ET 198
Holster yellow
120419
ET 247
Test Lead Set
120426
* To ensure safety, use Bussmann BBS-2 only.