Model 182 Radon Flask Counter
Ludlum Measurements, Inc.
Page 8
June 2011
Excessive Background Count On Plateau
A. Expected Background Count
3 counts per 10 minutes or less with
scintillation material
B. Excessive: Light leak
Cable malfunction
Microswitch malfunction
PMT noisy
Instrument contaminated