0 Scaler count time = 30
seconds
0 Display units = counts
o
Display time base minutes
0 Display range multiplier =
auto
o Detector dead time
o Detector calibration
1
Detector model =
44-2
Detector serial number =
PRO73738
Ratemeter alarm setting =
Integrated dose alarm setting
overload current = off
RESPONSE
Select the parameters display
The
display should
now look like
4.
The high voltage should now
read 700v.
The threshold should now read
100.
The window should now be
Off.
The count time should now
The
display should
now look like
5.
units should now read 7.
‘The time base should now
read 1.
mode.
Set the high voltage to 700V.
Set the threshold to 100.
Tloo
Turn the window
Set the scaler
time to 30
seconds.
Select the detector display mode.
set the readout units to counts.
Set the readout range multiplier
Set the readout
time
base to
Set the detector dead time to
Set the
constant to 1.
Set the model to
44-2.
Set the serial number to
Select the alarm display mode.
Set the ratemeter alarm to
Turn the overload
The dead time should now
read
The range multiplier should
The calibration constant
should mad 1.
now read 0.
model should read
44-2.
The serial number should now
read PR073738.
The
display should
now look like Figure 6.
‘The
alarm should
now read
scaler alarm should now
‘The integrated dose alarm
should now read
The overload should now be
page 22