Clear the logging memory to SSC
allow for new data.
logging
should
be cleared and the sample
number should be 0.
Connect the
44-2 and then
integrated dose and the
clear the integrated dose.
integrated dose timer should
he 0.
RESPONSE
Move the north side of the
N
room and enter the location code.
Log the scaler reading.
The location code should
change to OUTSIDE N.
The scaler will start to count.
Remain at the same position
until the count is completed.
When
scaler stops coun-
ting the reading is saved as
sample 0 and the sample
counter increments to 1.
Move to the east side of the
room and enter the location code.
Log the scaler reading.
Move to the south side of the
S
room and enter the location code.
Log the
reading.
Move to the west side of the
W
room and enter the location
location code should
change to OUTSIDE E.
Remain at the same position
until the count is
When the scaler stops count-
ing, the reading is saved as
sample 1 and the sample
counter increments to 2.
location code should
change to OUTSIDE S.
The scaler will start to count.
Remain at the same position
until the count is
When the scaler stops count-
ing the reading is saved as
sample 2 and the sample
location code should
change to OUTSIDE W.
The scaler will start to count.
Remain at the same position
until the count is completed
When the scaler stops count-
ing the reading is saved as
sample 3 and the sample
page28