Power Quality Analyzer User’s Guide
14
Place the update files under the root directory of U disk, then insert to the Analyzer.
After the U-disk is recognized by device, enter into user configuration interface from
main menu, press
【
F1
】
to enter into update interface.
There are tree types of update file including Software, Firmware and System. Select
you wanted files to update, the Analyzer will give prompt after finishing update, then
plug out the U disk, power off the Analyzer then power on to complete the update.
Error codes in right of below table are possible to pop up during update process,
coming with the solution in the left area.
Error code
Solution
"ErrCode: 0000 XXXX"
May be Flash damage, please replace Flash.
"ErrCode: 0001 XXXX"
SPI FLASH model are not supported, please
check the update file.
"ErrCode: 0003"
Error Verify of firmware data, please check the
update file.
"ErrCode: 0005 XXXX"
Abnormal status, please feedback XXXX to
manufacturer.
"ErrCode: 0010"
Serious overtime. Don’t outage and power off.
Please firmware update again.
"ErrCode: 0011"
Please try to firmware update again.
XXXX means the detail error message,please feedback to manufacturer.