NP45-09 User’s Guide
29
Bar Graph Screen
The Bar Graph display shows the percentage contribution of each of the components
related to the fundamental or full signal. A signal without distortion should show a 1
st
harmonic at 100% while the others are at 0: in practice this will not occur because
there always is a certain amount of harmonics resulting in distortion.
A sinewave becomes distorted when harmonics components are added to it. Distortion
is represented by the Total Harmonic Distortion (THD) percentage. The display can
also show the percentage of the DC component and each harmonic ratio.
【
◄
】【
►
】
arrow keys are used to position the Cursor on particular bar. The screen
header will show harmonic voltage/current, harmonic component ratio, frequency and
phase angle. If not all bars are shown on the screen, you can bring the next set within
the viewing area by moving the Cursor off the left or right end of the screen.
Function keys instruction:
【
F1
】
: Selection of harmonics type: voltage, current.
【
F2
】
: Selection of bar set to be displayed: L1, L2, L3, N or all
【
F3
】
: Interharmonics display on/off
【
F4
】
: Acess to table screen
【
F5
】