P30P-09A 83 User's manual
displayed in cycles.
Err Cal.
No calibration parameters - send the transducer for maintenance, message
does not block the display of measured values – it is displayed in cycles.
Err Batt
Service
Real Time Clock battery voltage too low – RTC settings will be deleted after the
transducer is powered off, transducer can operate, it is recommended to send
the transducer for battery change; message does not block the display of
measured values – it is displayed in cycles, setting of date or time turns the
message off.
Err Par.
Parameters error – transducer settings error, return to factory defaults,
transducer can operate but it is not recommended until the factory settings are
reverted, message does not block the display of measured values – it is
displayed in cycles
Error file
Attempt to read the file from external SD/SDHC or file system internal memory
failed – no file present, or invalid file format, transducer can operate, message
does not block the display of measured values – it is cycled approx. every 20
seconds.
8. Technical data
Input:
Table 50
Measuring value
Measuring range
K
U
=1, K
I
=1
Class (1s)
RMS current
I, I
DM
1 A
5 A
0.01 ...1...1.200 A~
0.05 ...5... 6.000 A~
0.2 %
RMS voltage
U
,
100V
(depends on ordering code)
230V
6... 100...120 V
12.5 ..230.. 300 V
0.2 %
Frequency
f
2....40.0 .. 60.0 .. 100 Hz
0.1 %
Active power
P
[W]
Reactive power
Q
[var]
1A, 100V
-144..-100 ... 100..144
0.5 % [W]
0.5 % [var]
5A, 100V
-720..-500 ... 500..720
1A, 230V
-360..-230 ... 230..360
5A, 230V
-1800..-1150 ... 1150..1800
Apparent power
S
1A, 100V
0 ... 100..144
0.5 % [VA]
5A, 100V
0 ... 500..720