2000-T-INS-LAB-RevC14 

13

TROUBLESHOOTING

If  the device does not function properly:
1.  Ensure the battery is properly installed or replace the battery. Be sure 

to observe proper polarity markings when replacing the battery. If  the 
yellow indicator light on the front of  the unit no longer illuminates when 
it is turned on, replace the battery and check again.

2.  If  the indicator light is flashing, the intensity has been adjusted, and no 

stimulation is felt, check to ensure the lead wires are properly connected 
and the electrodes are properly applied to the skin. If  the unit appears to 
be functioning and no stimulation is felt, the lead wires or electrodes may 
need to be replaced.

3.  If  the battery appears to be charged and the unit is not functioning, turn 

both Intensity Control Knobs to the OFF position (counterclockwise). 

Then gradually turn the Intensity Control Knob (clockwise) until 
stimulation is felt. If  device still is not working, turn the unit off  and 
contact your authorized GF Health Products, Inc. distributor.

If  there is any other problem, please contact an authorized GF Health 
Products, Inc. distributor. Do not try to repair a defective device.

Summary of Contents for 2000-T

Page 1: ...Stimulation TENS Model 2000 T Operation Manual Read this manual before operating your 2000 T Save this manual for future use The most current version of this manual can be found online at www grahamf...

Page 2: ...NFORMATION 11 CARING FOR YOUR TENS DEVICE 12 TROUBLESHOOTING 13 TECHNICAL SPECIFICATIONS 14 OUTPUT SPECIFICATIONS 14 LIMITED WARRANTY 15 GF Health Products Inc is not responsible for typographical err...

Page 3: ...nate the pain sensation felt by masking the original pain message sent to the brain It is also believed that TENS stimulation helps release endorphins into the blood stream thereby further reducing pa...

Page 4: ...situation or unsafe practice that if not avoided could result in death or serious personal injury s CAUTION Indicates a potential hazard or unsafe practice that if not avoided could result in minor p...

Page 5: ...ive mechanism WARNING Notice for California Customers California Proposition 65 WARNING This product contains a chemical known to the State of California to cause cancer and reproductive or developmen...

Page 6: ...ols are very easy to use and the slide cover protects accidental changes in settings MODE uS PULSE WIDTH Hz PULSE RATE Lead Connector Intensity Controls Indicator Light Pulse Width Control Mode Select...

Page 7: ...idth knob regulates the pulse width for both channels Pulse Rate The Pulse Rate knob regulates the number of pulses per second for both channels Mode Functions Burst B releases individual bursts twice...

Page 8: ...s located on top of the unit Holding the insulated portion of the connector push the plug end of the wire into one of the ports one or two sets of the wires may be used After connecting the wires to t...

Page 9: ...mended by your physician clinician apply the skin treatment as recommended let dry and apply electrodes as directed Following these recommendations will both reduce the chance of skin irritation and e...

Page 10: ...recommended by your physician clinician 3 Insert lead wire connector into the device plug end of lead wire into the channel output port jack to be used push the plug in as far as it will go 4 Select...

Page 11: ...yellow indicator light on the front of the unit no longer illuminates the battery should be replaced 1 Turn unit OFF 2 Remove the front panel cover by pressing on the top of the panel and pressing dow...

Page 12: ...p and water Do not immerse the device in water or other liquids Wipe lead wires with damp cloth moistened with soap and water Do not immerse the lead wires To properly store the device for an extended...

Page 13: ...d wires are properly connected and the electrodes are properly applied to the skin If the unit appears to be functioning and no stimulation is felt the lead wires or electrodes may need to be replaced...

Page 14: ...ave Form Asymmetrical bi phasic square pulse Voltage 0 100 Volt open current Power Source 9 Volt battery Dimensions 95 mm x 65 mm x 23 5 height x width x thickness Weight 115 g including battery OUTPU...

Page 15: ...368 2386 or e mail cs grahamfield com Failure to follow the specific directions provided by the GF customer service team will result in denial of the warranty claim ENTIRE WARRANTY EXCLUSIVE REMEDY AN...

Page 16: ...Manufactured for GF Health Products Inc 2935 Northeast Parkway Atlanta Georgia 30360 USA tel 770 368 4700 fax 770 368 2386 www grahamfield com 2008 GF Health Products Inc Made in Taiwan...