Tests...Operating principle
6-156
LTO
6.2.2 O
PERATION
OF
THE
OTHER
SCREENS
6.2.2.1 P
RESENTATION
The following screen corresponds to the solenoid valve and pump tests:
Screen 6-2: Test screen
6.2.2.1.1
S
CREEN
WITH
SWITCH
To start the test, press the switch of the element you wish to test.
Observe the test.
To stop the test, press the switch again.
6.2.2.1.2
S
CREEN
WITH
MICRO
-
SWITCHES
AND
PHOTOCELLS
Manually operate the micro-switch or photocell of the element you wish to test.
Observe the status of the micro-switch or photocell on the screen.
Spraying
of edgers
and visor
Spraying
Rear of edging
chamber
Start-up of
edgers
(visor closed)
Pump
Exit
Micro-
switches