6-4.ALARM function:
6-4-1.Upper and lower range of settings: eg 4-5-7 Option Description
6-4-2. After the range setting, press
▼
/ ALARM KEY> 2 SEC. ALARM function is
entered, and the top right corner of the screen will appear the words ALM.
(eg SCREEN16)
6-4-3.When the measured value equivalent
≧
upper limit set value or
≦
lower
limit value, and the digital will flash a warning sound, there will be alarm
output (Open collector)(3-11, Fig.1).
A.Hysteresis = Measured value * 5 %.
B. High ALARM status:
Measured value
≦
(upper limit set value - Hysteresis value), the Alarm
goes OFF.
C. Low ALARM status:
Measured value
≧
(lower limit value + Hysteresis value), the Alarm goes
OFF.
D.Example: PM upper limit set value=50, PM lower limit value=30.
D-1.When the measured value PM=50, then ALARM to ON.
D-2.When the PM when the measured value=45, H=50 * 5%=3,
(45
≦
(50-3)), so 45
≦
47 ALARM goes OFF.
D-3.When the measured value PM=30, then ALARM to ON.
D-4.When the measured value PM=35, H=30 * 5% =2, (35
≧
(30+2)),
so 35
≧
32 ALARM goes OFF.
6-4-4.Press
▼
/ ALARM KEY> 2 SEC. ALARM function is canceled.
(eg SCREEN1)
6-4-5. ALARM function interface Description.
11