9. Air filter net REPLACEMENT
When the meter show " " information ,
This " symbol " meaning is please replace the meter filter,
Please carry out according to the following replacement steps.
1) Loose the screws of the " Air filter net Cover " ( 3-14, Fig. 1 )
and take away the " Air filter net Cover " from the instrument
and remove the Air filter net.
2) Replace with Air filter net and reinstate the cover.
3) Make sure the cover is secured after changing Air filter net.
4) please power off the meter ,press and hold the HOLD key & REC key then
power on the meter, at the same time the meter will be show power on screen
and one down counter and " Replace Time Clear " text , when the down counter
decries to zero, the " Replace Filter " Information will be remove.
10. POWER SUPPLY from DC
ADAPTER
The meter also can supply the power supply from the DC 9V
Power Adapter . Insert the plug of Power Adapter into
" DC 9V Power Adapter Input Socket "( 3-13, Fig. 1 ),
Then Press and hold " Power Button"( 3-2, Fig. 1 ) > 2 sec.
The meter will be power ON .
11. BATTERY REPLACEMENT
1) When the left corner of LCD display show " ", it
is necessary to replace the battery. However, in-spec.
measurement may still be made for several hours after
low battery indicator appears before the instrument
become inaccurate.
2) Loose the screws of the " Battery Cover " ( 3-16, Fig. 1 )
and take away the " Battery Cover " from the instrument
and remove the battery.
3) Replace with DC 1.5 V battery ( UM3, AA,
Alkaline / heavy duty ) x 6 PCs, and reinstate the cover.
4) Make sure the battery cover is secured after changing batteries.
21
Summary of Contents for PM-1064SD
Page 22: ...20 ...