Standard value
1.318
1.413
1.318
1.318
* If only intend to make the one point calibration,
just execute the 2 mS range ( 1.413 mS Cal. )
is enough.
* Multi-points calibration procedures should
execute the 2 mS range ( 1.413 mS Cal. )
calibration at first, then make other ranges
(20 uS range, 20 mS range or 200 mS range )
calibration procedures following if necessary.
7. SALT MEASURING and
CALIBRATION
23