GB
20
1.2.1 Electrical data
Internal power supply
4.5 V (DC); Approved battery types
for the power supply:
• VARTA Alkaline, No.4001, Size LR1, 1.5V
• VARTA High Energy, No.4901, Size LR1, 1.5V
• DURACELL Alkaline, Type MN9100, Size N, 1.5V
• Energizer Alkaline; Type E90, LR1, 1.5V
Change only outside of hazardous location.
Module input
Explosion protection class intrinsic safety EEx ib IIB;
[Socket 1(+), 3(-), 4, 6
or
Max. values:
internal plug St1 or St2]
U
o
=
5 V
I
o
=
55 mA
P
o
=
68 mW
Characteristic curve: linear
C
i
=
1.1 µF
L
i
unconsidered small
Max. admissible outer values for:
(following values correspond to the computer program
for PTB report ThEx-10)
L
o
(mH)
IIB
C
o
(µF)
2
16
1
19
0.5
24
when connecting with an active module output in explosion
protection class intrinsic safety with max. admissible values:
U
o
=
6.6 V
I
o
=
165 mA
P
o
=
272 mW
Characteristic curve: linear
following max. admissible outer values are valid for the connection
with the module: (these values correspond to the computer
program for PTB report ThEx-10)
L
o
(mH)
IIB
C
o
(µF)
2
6.9
1
9.9
0.5
12