Scenario 2 — Dual chillers
Separate chillers are provided for the laser resonator/DC power supply and for the AOM module.
2
1
3
4
5
1
Coolant in — laser/DC power supply
2
Coolant out — laser DC power supply
3
DC power supply
4
Coolant in — AOM module
5
Coolant out — AOM module
Figure 16 - Dual chiller coolant connections
Table 16 - Coolant flow requirements
Flow (L/min)
SR 10 AOM
SR 25 AOM
Laser/DC power supply minimum flow
≥ 4
≥ 5
Laser/DC power supply recommended flow
≥ 5
≥ 6
AOM module minimum flow
≥ 2
≥ 2
AOM module recommended flow
≥ 3
≥ 3
CAUTION
Risk of damage to optical components within the AOM module
The coolant pressure within the AOM module must not exceed 3bar. Higher pressures may
cause leakage within the module resulting in irreversible damage to optical components.
An inline pressure regulator, set to 3bar, is supplied with the system and must be fitted to the
coolant inlet pipe as shown in the following figure. The pressure regulator is directional and must be
fitted with the arrow in the same sense as the coolant flow (pointing toward the inlet connector).