Description of the incubator
:
A
- distilled water tank - 4L.
B-
water level indicator
C-
air holes
D-
controller
E-
incubator chamber
F-
transparent door to control the raising process
G-
incubator door
Figure 2
A B C D E
Description of the controller:
A-
main switch 0-1
B-
controlling panel
C-
power supply cable
D-
5A fuse holder
E-
Bluetooth communication outlet
2.2 Handing the controller
ŁYSOŃ IC-0x 0.1
. . . . . . .
2
3
4
5
1
ELEMENT
DESCRIPTION
1
–
display
After
activation
–
displays
sequentially: manufacturer’s name,
device type and software version
1. Main screen
–
basic screen
selected by default once the device
has
been
activated.
Screen
description below.
2. The screen to set the setpoint
temperature
–
the displays shows
the temperature measured (upper
line) and the setpoint temperature
(lower line).
3. The screen to set the setpoint
humidity
–
the display shows the
humidity measured (upper line) and
the setpoint humidity (lower line).
4. The screen for operating
statistics
–
the displays shows the
minimal and maximum values of
temperature and humidity registered
(since the control supply activation).
Reset of the values registered shall
occur after pressing the „plus” or
„minus” buttons
5. Service screen for TH sensor
temperature readouts calibration
–
the display shows the temperature
measured including the calibration
(upper line) and the calibration
value (lower line)
6. Service screen for humidity
readouts calibration
–
the display
shows the humidity measured
including the calibration (upper
line) and the calibration value
(lower line)
7. Service screen for temperature
regulator tuning
–
the display
shows the value of Kp parameter
–
tuning of the temperature regulator.
In
case
when
the
setpoint
temperature has not been achieved
or when reaching the temperature
lasts too long, the value of the
parameter must be increased.
Respectively, when there are