noticeable
temperature
over-
regulations, the parameter shall be
lowered.
8.
Service
screen
for
the
calibration of T1 additional
sensor temperature readouts.
The
display shows the temperature
measured including the calibration
(upper line) and the calibration
value (lower line)
2
–
“
-
“ button for
value lowering
Main screen (1)
–
pressing the
button will switch off the lighting of
the incubator chamber.
Statistics screen (4)
–
pressing the
button will reset the statistics for the
measurements of the temperature
and humidity.
Other screens
–
pressing the button
will decrease the value of the
modified parameter.
3
–
„+„
button
for value
increasing
Main screen (1)
–
pressing the
button will switch on the lighting of
the incubator chamber.
Statistics screen (4)
–
pressing the
button will reset the statistics for the
measurements of the temperature
and humidity.
Other screens
–
pressing the button
will increase the value of the
modified parameter.
4
–
button
„<>”
screen switching
Pressing the button shortly will
switch between the controller
screens.
5
–
button
„ON/OFF”
Pressing the button alternately for a
short time will activate (ON) and
deactivate (OFF) the climate
controlling inside the incubator.
In the deactivated state (OFF), the
regulator
acts
as
a
typical
hytherograph completed by the
registration system for min and max
values.
Additionally,
in
the
deactivated state it is possible to
switch on and off the cabin lighting,
switch between the screens, reset
the operating statistics and to
modify the parameters displayed.
In the activated state (ON), apart
from all the operations typical for
OFF
state,
the
controller
additionally :
•
based on the TH sensor
readout, maintains the
setpoint temperature by
regulating
the
power
supply to the main heater.
•
based on the TH sensor
readout,
maintains
the
setpoint humidity
•
based on the T1 sensor
readout,
maintains
the
setpoint temperature by
regulating
the
power
supply to an additional
heater
•
regulates the water level
inside
the
humidifier
system
•
detects
and
signals
emergency and warning
states
26.5 C
43%
T D W
1
2
3
4
6
8
9
10
7
5
ELEMENT
DESCRIPTIONS
1
–
temperature
indication
The field shows the value of the
temperature measured.
2
–
signalling of
T1 sensor damage
Displayed sign together with a
periodical sound signal indicate
the damage to an additional sensor
(T1). Switching off the controller
resets the failure signalling.
3
–
signalling of
open door
Displayed sign together with a
periodical sound signal indicate
timeout of door opening. Closing
the door resets the signal
4
–
signalling a
failure in the
water filling
system of a
humidifier
Displayed sign together with a
periodical sound signal indicate
the timeout for water filling in a
humidifier. The signal may be
triggered by:
•
water shortage
•
damage/suspension of a
float switch