background image

 

noticeable 

temperature 

over-

regulations,  the  parameter  shall  be 
lowered.  

8. 

Service 

screen 

for 

the 

calibration  of    T1  additional 
sensor  temperature  readouts. 

The 

display  shows  the  temperature 
measured  including  the  calibration 
(upper  line)  and  the  calibration 
value (lower line) 

 

-

“ button for 

value lowering 

 

Main  screen  (1)

 

  pressing  the 

button will switch off the lighting of 
the incubator chamber.  

Statistics  screen  (4)

 

  pressing  the 

button will reset the statistics for the 
measurements  of  the  temperature 
and humidity.  

Other screens

 

 pressing the button 

will  decrease  the  value  of  the 
modified parameter.  
 

  

„+„

 button 

for value 

increasing 

 
Main  screen  (1)

 

  pressing  the 

button will switch on the lighting of 
the incubator chamber.  

Statistics  screen  (4)

 

  pressing  the 

button will reset the statistics for the 
measurements  of  the  temperature 
and humidity.  

Other screens

 

 pressing the button 

will  increase  the  value  of  the 
modified parameter.  

 button 

„<>” 

screen switching 

 
Pressing  the  button  shortly  will 
switch  between    the  controller 
screens. 

 

 button 

„ON/OFF”

 

Pressing the button alternately for a 
short  time  will  activate  (ON)  and 
deactivate  (OFF)  the  climate 
controlling inside the incubator. 
 
In  the  deactivated  state  (OFF),  the 
regulator 

acts 

as 

typical 

hytherograph    completed  by  the 
registration system for min and max 
values. 

Additionally, 

in 

the 

deactivated  state  it  is  possible  to 
switch on and off the cabin lighting, 
switch  between  the  screens,  reset 
the  operating  statistics  and  to 
modify the parameters displayed.  
 
In  the  activated  state  (ON),  apart 
from  all  the  operations  typical  for 
OFF 

state, 

the 

controller 

additionally : 
 

 

based  on  the    TH  sensor 

readout,    maintains  the 
setpoint  temperature  by 

regulating 

the 

power 

supply to the main heater. 

 

based  on  the  TH  sensor 

readout, 

maintains 

the 

setpoint humidity 

 

based  on  the  T1  sensor 

readout, 

maintains 

the 

setpoint  temperature  by 
regulating 

the 

power 

supply  to  an  additional 
heater 

 

regulates  the  water  level 

inside 

the 

humidifier 

system 

 

detects 

and 

signals 

emergency  and  warning 
states 

 

 
 
 
 
 
 
 

      26.5  C
      43%

T   D  W

1

2

3

4

6

8

9

10

7

5

 

 

ELEMENT 

DESCRIPTIONS 

 temperature 

indication 

 
The  field  shows  the  value  of  the 
temperature measured. 
 

 signalling of 

T1 sensor damage 

Displayed  sign  together  with  a 
periodical  sound  signal  indicate  
the damage to an additional sensor 
(T1).  Switching  off  the  controller 
resets the failure signalling.  
 

 signalling of 
open door  

Displayed  sign  together  with  a 
periodical  sound  signal    indicate 
timeout  of  door  opening.  Closing 
the door  resets the signal 
 

 signalling a 

failure in the 

water filling 

system of a 

humidifier 

Displayed  sign  together  with  a 
periodical  sound  signal  indicate 
the  timeout    for  water  filling  in  a 
humidifier.    The  signal  may  be  
triggered by: 
 

 

water shortage 

 

damage/suspension  of  a 

float switch 

Summary of Contents for W5014

Page 1: ...si biorstwo Pszczelarskie Tomasz yso Sp ka z o o Sp ka Komandytowa 34 125 Su kowice ul Rac awicka 162 Polska www lyson com pl email lyson lyson com pl tel 33 875 99 40 33 870 64 02Siedziba Firmy Klecz...

Page 2: ...ety principles for device operation 1 1 Operational principles 1 2 Electrical safety 1 3 Operational safety 2 Characteristics of the device 2 1 Device description 2 2 Controller handling 2 3 Error cod...

Page 3: ...idden to pull the power supply cable The power supply cable must be kept away from any heat sources sharp edges and its proper state must be secured 1 3 Operational safety 1 The following equipment is...

Page 4: ...humidity the display shows the humidity measured upper line and the setpoint humidity lower line 4 The screen for operating statistics the displays shows the minimal and maximum values of temperature...

Page 5: ...for min and max values Additionally in the deactivated state it is possible to switch on and off the cabin lighting switch between the screens reset the operating statistics and to modify the paramet...

Page 6: ...ocked ON OFF button 300 Damaged sensor of temperature measurement for TH composite sensor 301 Temperature indication of TH composite sensor outsider the permissible range 302 Temperature regulation er...

Page 7: ...ectrical equipment distributor network at least free of charge and directly if the device to be returned is of proper type and serves the same purpose as the newly purchased device 6 Guarantee Product...

Reviews: