- 28 -
9.2.4.2
Fill a small amount (0,5 ml till maximum 2,0 ml) of distilled or sterile water into
the medication chamber and close the M-neb
®
dose
+
nebulizer unit MN-300/8,
by replacing the screw cap and closing it by rotating clockwise 80°.
9.2.4.3
After this, connect the filled M-neb
®
dose
+
nebulizer unit MN-300/8 with the M-
neb
®
dose
+
mesh nebulizer MN-300/8 (control unit). To do this, connect the
female USB socket of the control unit with the male USB connector of the
nebulizer unit.
9.2.4.4
Switch the M-neb
®
dose
+
mesh nebulizer MN-300/8 (control unit) on. For this,
press (> 1 second) the on/off button of the M-neb
®
dose
+
mesh nebulizer MN-
300/8 (control unit). The M-neb
®
dose
+
mesh nebulizer MN-300/8 switches on
automatically. As indication, you will hear a short acoustic signal tone and the
battery status indicator continuously lights up green.
The operational readiness of the M-neb
®
dose
+
mesh nebulizer MN-300/8
(control unit) is established after the operating mode status indicator has
flashed three times blue briefly.
The M-neb
®
dose
+
mesh nebulizer MN-300/8 is now in triggered mode.
9.2.4.5
Now you can change the operating mode from the triggered mode to continuous
mode in a time window of 60 seconds, which is equivalent to the maximum
standby time. For this, press (< 1 second) the on/off button of the M-neb
®
dose
+
mesh nebulizer MN-300/8 (control unit). Two consecutive, short, acoustic signal
tones sound and the operating mode status indicator lights up continuously blue.
9.2.4.6
The M-neb
®
dose
+
mesh nebulizer MN-300/8 is now in the continuous mode
and continuously nebulises the filled distilled or sterile water. Once the filled
distilled or sterile water has been completely nebulised, the auto off function
automatically switches off the M-neb
®
dose
+
mesh nebulizer MN-300/8.
WARNING
Do not inhale the aerosolised distilled or sterile water.
Ensure sufficient ventilation of the room in which you perform the cleaning.
9.2.4.7
Now disconnect the M-neb
®
dose
+
nebulizer unit MN-300/8 from the M-neb
®
dose
+
mesh nebulizer MN-300/8. Disconnect the screw cap of the M-neb
®
dose
+
nebulizer unit MN-300/8 and put the M-neb
®
dose
+
nebulizer unit MN-300/8 on
a dry, absorbent surface to dry.
9.2.4.8
After the M-neb
®
dose
+
nebulizer unit MN-300/8 has dried in the air, you can
reconnect the screw cap of the M-neb
®
dose
+
nebulizer unit MN-300/8.
Reconnect the M-neb
®
dose
+
mouthpiece MN-300/8 again with the M-neb
®
dose
+
nebulizer unit MN-300/8.
10.0 Maintenance
The M-neb
®
dose
+
mesh nebulizer MN-300/8 should be checked after a maximum of 24
months and subjected to a technical safety inspection. All repairs or maintenance may only
be done by a dealer qualified by NEBU-TEC med. Products Eike Kern GmbH.