20VS5-213
P. 4 / 4
EM-4429 Rev.1
For a non-inverting amplifica-
tion with external adjustments,
when the combined resistance of
the span adjustment and 9kΩ is
10kΩ, the gain is (1 + 10kΩ / R
1
)
×
V
IN
. With the span adjustment
(2kΩ), the output is configurable.
The zero is adjustable with a ref-
erence voltage source of ±15V DC
and the attenuation of 200Ω and
30kΩ.
For an inverting amplification
with external adjustments, when
the combined resistance of the
span adjustment and 9kΩ is
10kΩ, the gain is -(10kΩ / R
2
)
×
V
IN
. With the span adjustment
(2kΩ), the output is configurable.
The zero is adjustable with a ref-
erence voltage source of ±15V DC
and the attenuation of 200Ω and
30kΩ.
MAINTENANCE
Regular calibration procedure is explained below:
■
CALIBRATION
Warm up the unit for at least 10 minutes. Apply 0%, 25%,
50%, 75% and 100% input signal. Check that the output
signal for the respective input signal remains within accu-
racy described in the data sheet.
M-SYSTEM WARRANTY
M-System warrants such new M-System product which it manufactures to be free from defects in materials and workmanship during the 36-month period following the date that such
product was originally purchased if such product has been used under normal operating conditions and properly maintained, M-System’s sole liability, and purchaser’s exclusive
remedies, under this warranty are, at M-System’s option, the repair, replacement or refund of the purchase price of any M-System product which is defective under the terms of this
warranty. To submit a claim under this warranty, the purchaser must return, at its expense, the defective M-System product to the below address together with a copy of its original
sales invoice.
THIS IS THE ONLY WARRANTY APPLICABLE TO M-SYSTEM PRODUCT AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING ANY IMPLIED
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. M-SYSTEM SHALL HAVE NO LIABILITY FOR CONSEQUENTIAL, INCIDENTAL OR
SPECIAL DAMAGES OF ANY KIND WHATSOEVER.
M-System Co., Ltd., 5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN, Phone: (06) 6659-8201, Fax: (06) 6659-8510, E-mail: info@m-system.co.jp
+
–
+
–
Non-inverting amplifier circuit zero/span adjustments (input ±5V, output ±10V)
OUTPUT ±10V DC
■
Non-inverting amplifier's circuit with external adjustments: Example of G = 2
+
–
–
+
REFERENCE
VOLTAGE
SOURCE
9kΩ
10kΩ
50kΩ
(ZERO)
2kΩ
(SPAN)
INPUT ±5V DC
10kΩ
30kΩ
200Ω
+
–
1
2
4
3
5
6
+
–
EXTERNAL
CLOCK / POWER
0 - 15V
17
16
14
15
+
–
Inverting amplifier circuit zero/span adjustments (input ±10V, output ±10V)
■
Inverting amplifier's circuit with external adjustments: Example of G = -1 (output inverted to the input)
+
–
–
+
REFERENCE
VOLTAGE
SOURCE
9kΩ
10kΩ
100kΩ
(ZERO)
2kΩ
(SPAN)
INPUT ±10V DC
10kΩ
30kΩ
200Ω
+
–
+
–
OUTPUT ±10V DC
1
2
4
3
5
6
+
–
EXTERNAL
CLOCK / POWER
0 - 15V
17
16
14
15
■
Multi-channel
20VS5
–
2
20VS5
–
213
8ch
8
1
9
2
POWER SUPPLY
15
14
20VS5
–
213
・
・
・
・
・
・
15
14