M3LDY

EM-2664 Rev.2

P. 4 / 6

CHECKING

1) Terminal wiring:  Check that all cables are correctly connected according to the connection diagram.
2) DIP SW setting:  Check that the switches are set to appropriate positions.
3) Power input voltage:  Check voltage across the terminal 11 – 12 with a multimeter.
4) Input:  Check that the input signal is within 0 – 100% of full-scale.
5) Output:  Check that the load resistance meets the described specifications.

I/O RANGING & FINE ADJUSTMENTS

After the DIP SW setting is complete, set up the precise input and output range using the front control buttons.
The front LEDs’ colors and flashing patterns help you to easily identify the transmitter’s status and confirm the setup actions
in each step of Calibration Modes.  Please read the following explanations referring to “Calibration Flow Chart” below.

POWER ON

Amber LED

Green LED

Control
Buttons

Red LED

OFF

ON

Blink

RUN MODE

LD1

LD2

LD3

A

G

RUN MODE

LD1

LD2

LD3

G

R

LD1

LD2

LD3

LD1

LD2

LD3

0% INPUT

CONFIGURED

100% INPUT

CONFIGURED

INPUT

CALIBRATION

MODE

LD1

OUTPUT

CALIBRATION

MODE

LD2

LD3

LD1

LD2

LD3

0% OUTPUT

CONFIGURED

100% OUTPUT

CONFIGURED

G

G

R

R

R

R

G

R

R

G

Fine Adjustments ?

I/O Ranging ?

PUSH  

MODE

 

 1 – 2 s

PRESS & HOLD  

MODE

 

 > 5s

PUSH  

MODE

PUSH  

MODE

PUSH  

MODE

PUSH  

MODE

PUSH

DOWN

PUSH

UP

DOWN

UP

MODE

PUSH

DOWN

PUSH

UP

When you set 0% or 100% input/output ranges, 
keep pressing UP or DOWN button until the
LD1 flashes for approx. 2 seconds and turns off, 
which indicates the setup is complete.
When you release the button, the LD1 is returned
to ON.
If the LED does not change, the entered level may
be inappropriate, i.e. out of usable range.

FINE ZERO

CALIBRATION MODE

UP

  or  

DOWN

FINE SPAN

CALIBRATION MODE

UP

  or  

DOWN

Apply simulated 0% input signal.

Apply simulated 100% input signal.

Adjust simulated input until the output meter 
shows desired 0% output.

Adjust simulated input until the output meter 
shows desired 100% output.

*

1. Fine zero and calibrations are performed for 0% and 100% output regardless of the input value.

*

1

*

1

■ 

CALIBRATION FLOW CHART