• Follow all local safety codes before entering the vessel.
1. Remove any foreign magnetic particles from the mixing head.
If the mixing head must be placed on a table, place the female bearing downward to
prevent magnetic particles from adhering to the magnetic surface of the mixing head.
2. Remove the drive unit.
3. Install the mixing head using the multi-tool. Mixing head sizes GMP 5 000–30 000
must be installed by hand.
4. Carefully position the mixing head by turning it while lowering it onto the male
bearing.
5. Align the mixing head with the tank plate. Ensure that the mixing head rotates
smoothly.
Drive Unit Installation
• The drive unit may be heavy. Use appropriate equipment to avoid injury.
• Never use the drive unit without a frequency converter. See Control Cabinet in this
guide.
• See the Electrical Installation section of this manual for electrical information.
Check direction of rotation
Turn on the motor. Verify that the drive head rotates clockwise and that the fan rotates
counter clockwise. A sticker on the motor indicates the correct rotation direction of the
fan.
!