Mounting the Drive Unit
• Handle the drive unit with caution to avoid personal injury and damage to the
equipment.
• Do not make any adjustments to screws and bolts.
• There is a powerful magnetic force between the mixing head and the outer drive
head.
Mount the drive unit onto the vessel following the instructions for the appropriate mixer
size:
GMP 50, GMP 100, GMP 500, and GMP 50 DC
1. Install the drive unit into the tank plate. Put the two studs on the tank plate into the
slots on the drive unit.
2. Turn the drive unit clockwise (view from below) to the stop position, approximately
15º.
3. Tighten the two nuts with a wrench.
15°
GMP 50,
100, 500
GMP 50,100, 500
GMP 1000 and GMP 2000
10°
GMP 1000, 2000
1. Tighten the screws (1) in a clockwise direction (view from below).
1
2. Install the drive unit into the tank plate. Insert the locking screws into the two oval
holes in the tank plate (2).
3. Turn the drive unit (3) clockwise (view from below) to the stop position, approximately 10º.
!
NovAseptic
®
GMP Mixer Installation and User Guide
13