Value at 4mA
The value of the “
Source
” channel at which the “
4-20mA output
” will be set to 4mA.
Value at 20mA
The value of the “
Source
” channel at which the “
4-20mA output
” will be set to 20mA.
Set at log time only
The user can elect to output an averaged value on the “
4-20mA output
” by checking the box. When checked,
the “
4-20mA output
” will only be updated at the logging interval otherwise it is updated at each
measurement interval. In the case of the measurement interval being shorter than the logging interval, then
the “
4-20mA output
” will be the average of those values calculated per measurement interval.
Click “
Apply
” to save the settings then click “
Exit
” to return to the “
Configure outputs
” dialogue box.
Test a "4-20mA" output
From the "
Edit 4..20mA output
" dialogue box, click "
Test
".
1. Enter the "
Output value
" that you wish to send to the connected "
4-20mA receiving device
".
2. Click "
Send
" and the set "
Output value
" will be sent.
Alternatively
:
From the "
Edit 4..20mA output
" dialogue box, click "
Test
".
1. Check the "
Scaled value
" checkbox and the "
Output value
" will change to output the current
proportional to the "
Channel value
". In this example, "
Velocity
".
MACE XCi User Manual - 230 -
Summary of Contents for HydroMace XCi
Page 1: ......
Page 13: ...MACE XCi User Manual 13...
Page 19: ...MACE XCi User Manual 19...
Page 33: ...MACE XCi User Manual 33...
Page 36: ...MACE XCi User Manual 36...
Page 167: ...NOTE None of these parameters can be edited MACE XCi User Manual 167...
Page 182: ...MACE XCi User Manual 182...
Page 187: ...MACE XCi User Manual 187...
Page 265: ...MACE XCi User Manual 265...