My XCi has problems measuring the depth
Symptom
Cause
Remedy
Depth readings respond
slowly to changes in actual
depth
Reference filter has not been
"
Enabled
"
"
Enable
" reference filter
Reference line has been
crimped
Check reference line for crimps
Sensor cable has been
squashed
Look for possible signs of cable
compaction
Depth readings seem to be
inaccurate
Incorrect units
(US only)
Check that units are in your
Slope/offset incorrectly entered
Check slope/offset entered cor-
rectly
Slope/offset units wrong
Check that slope/offset units
are correct off the tag
Slope/offset calculated incor-
rectly
Check calculation of slope/off-
Depth does not change
Damaged depth sensor
Check for damage and replace
ceramic module
Depth reading is always the
offset
Depth sensor not connected
properly
Check connection inside XCi
Damaged depth sensor
Check for damage and replace
ceramic module
Depth sensor stays at full-
scale
Damaged depth sensor
Check for damage and replace
ceramic module