Powering the XCi device with an external battery
In certain applications, it may be desirable to power the XCi device with an external battery. For example, if
the XCi device is connected to multiple sensors and is configured such that the measuring interval is short.
WARNING: MACE highly recommends that, if an external battery is used, that the
main internal XCi battery be left installed. The XCi battery not only provides some
"redundancy" in the system, but also provides a power source when the external
battery is removed during servicing.
1. The HydroMACE XCi devicecan be powered by an external 12V battery through the dedicated
“External Battery 12V” terminal.
NOTE: If a solar panel is connected to the
, the internal charging circuit
of the device will also charge the external battery. The charging circuitry of the
XCi device has a 20 Watt capacity.
WARNING: If the external battery is charged via its own solar panel and NOT
through the XCi charging circuit, the internal device battery will discharge
completely