12
Note:
• If the readingis<37.5°C(99.5°F ), the display will show together with green
LED.
• If the reading is
≥
37.5°C(99.5°F)and<43°C(109.4°F), the display will show
together with red LED and ten short beeps.
• Only after the unit °C/°F start to flash again, it can continue to take measu-
rement.
• As the forehead measurement temperature is likely to be affected by sweat,
oil and the surroundings, the reading shall be taken as a reference only.
• If the probe is placed at an angle close to the forehead measurement, the
reading will be affected by surrounding temperature.
• Babies’ skin reacts very quickly in the ambient temperature. Therefore, do
not take their temperature with the non-contact thermometer during/after
breastfeeding,because the skin temperature maybe lower than the internal
body temperature.
As a Object Thermometer
1.
Press the START/STOP button, all symbols appear
on the display , you can hear 1 short beep, then unit
°C / °F on the display start to flash.
2.
Press the SET button to select Object Mode, the
temperature unit flashes.
3.
Move the probe close to the object, make sure the
probe is flat and take measurement with a distance
between 1~3cm. Press the START/STOP button, The
meter will take measurement.
4.
The measurement will be finished in 1 second,
when it has been completed, you can hear 1 short
beeps, in the meantime, the reading will appears on
the display with green LED blinks.
Note:
• As the object scan temperature is likely to be affected by the surroundings,
the reading shall be taken as a reference only.
• If the probe is placed at an angle close the object scan, the reading will be
affected by surrounding temperature.